In this episode Shahriar takes a close look at a faulty Agilent E4433B 4.0GHz signal generator and vector modulator. This instrument is equipped with many of the most useful options such as dual arbitrary waveform generator and high-power output.
The instrument does not generate any error messages, however the output signal is ~20dB below specifications. After presenting the system block diagram and observing the output power characteristics, it becomes clear that the issue is with the mechanical attenuator. The teardown of the attenuator reveals faulty o-rings on the 5dB attenuation pad which has to be repaired with some flexible epoxy.
After the repair the instrument’s functionality is verified by measuring the output power with a power meter, frequency accuracy with a Rubidium frequency counter and the modulation accuracy with a vector signal analyzer.
Hitachi/Metabo electric can be purchased here:
The Signal Path
www.TheSignalPath.com
www.YouTube.com/TheSignalPath
www.Patreon.com/TheSignalPath
Rcreviews.lt is not the owner of this text/video/image/photo content, the real source of content is Youtube.com and user declared in this page publication as Youtube.com user, if you have any question about video removal, what was shared by open community, please contact Youtube.com directly or report bad/not working video links directly to video owner on Youtube.com. Removed video from Youtube.com will also be removed from here.

All Rights Reserved and belong to their respective owners. (Logos, Photos, Videos and etc.) Powered by MiReina.lt